S. No | Requirements | Surgical luminaire | |
Minor treatment | Major and system | ||
1 | Sterile handle( standard) | Yes | Yes |
2 | Central illuminance (Ec) | 40,000 - 160,000 | 40,000 - 160,000 |
3 | Light field diameter (d10) | Test value required | Test value required |
4 | Light distribution | Test value required | Test value required |
5 | Shadow dilution | Test value required | Test value required |
6 | Colour temperature (Tc) | 3000 - 6700 kelvin | 3000 - 6700 kelvin |
7 | Colour rendering index | 85 - 100 | 85 - 100 |
8 | Maximum value for total irradiance Ee | Test value required | Test value required |