| N | Measurand (measurable value) | Object of measuring | Method | Measuring range | Uncertainty |
| 1 | Absorbed dose rate to water | Calibrated field from linear accelerator | Measurements with chemical dosimeters | 1 × 102 ... 1 × 105 Gy∙s−1 | 2.5% |
| 2 | Absorbed dose to water | Chemical dosimeters | Irradiation in calibrated field from linear accelerator (high level) | 1 × 101 ... 1 × 106 Gy | 2.5% |
| 3 | Absorbed dose rate to water | Calibrated field from Sr-90/Y-90 radionuclide sources | Measurements with chemical dosimeters | 3 × 10−1 ... 1 × 102 Gy∙s−1 | 2.5% |
| 4 | Absorbed dose to water | Chemical dosimeters | Irradiation in calibrated field from Co-60 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.2% |
| 5 | Absorbed dose to water | Chemical dosimeters | Irradiation in calibrated field from Cs-137 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.2% |
| 6 | Absorbed dose to graphite | Chemical dosimeters | Irradiation in calibrated field from Co-60 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.2% |
| 7 | Absorbed dose to graphite | Chemical dosimeters | Irradiation in calibrated field from Cs-137 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.2% |
| 8 | Absorbed dose to aluminium | Chemical dosimeters | Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.5% |
| 9 | Absorbed dose to graphite | Chemical dosimeters | Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.5% |
| 10 | Absorbed dose to PMMA | Chemical dosimeters | Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 2.5% |
| 11 | Absorbed dose to polystyrene | Chemical dosimeters | Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 2.5% |
| 12 | Absorbed dose to polycarbonate | Chemical dosimeters | Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 2.5% |
| 13 | Absorbed dose to polystyrene | Chemical dosimeters | Irradiation in calibrated field from Co-60 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.2% |
| 14 | Absorbed dose to polystyrene | Chemical dosimeters | Irradiation in calibrated field from Cs-137 radionuclide sources (high level) | 1 × 101 ... 1 × 106 Gy | 1.2% |