2017 instrument conditions

ICP System, quartz glass “torch”

Agilent 7700x ICP-MS

RF forward power

1550 W

Frequency

27.12 MHz

Depth sampled

8 mm above load coil

Plasma torch Ar gas flow rate

15 L・min−1

He collision cell, ORS3

4.5 mL/minute He when ON

Kinetic energy discrimination of ORS3

4 V ED

Autosampler

ASX-500

Sample uptake rate

0.8 ml/min

Peristaltic pump

0.1 rps

Nebulizer

Concentric GE MicroMist

Nebulizer (carrier) Ar gas flow

1.0 L・min−1

Auxiliary (make-up) gas flow

0.1 L・min−1

Spray chamber

Scott double pass, quartz glass,

Peltier pump, 2˚C

Ni sampler cone orifice

1.0 mm

Ni skimmer cone orifice

0.7 mm

External-1 lens voltage

0 V (to improve stability)

Detector

Electron multiplier

Data collection; 3 scans through mass range per sample

Analyte peak jump mode

Total acquisition time per sample excluding wash

75.56 s

Each mass, 3 points 0.25 amu apart; Dwell time 0.1 s

0.3 s

Peak width at half height

~0.7 amu

CeO/Ce and Ce2+/Ce+

1.35% and 1.32%

2018 data collection differences

Data collection; 10 scans through mass range per sample

Analyte peak jump mode

Total acquisition time per sample excluding wash

75.56 s

CeO/Ce and Ce2+/Ce+

0.45(5)% and 0.97(9)%