Spectrometer Parameters | |||||
Hg Wavelength: 194.168 nm | Replicates 3 times | ||||
Integration time: 0.05 s | Transient Read Delay 0.0 s | ||||
Plasma parameters | |||||
Plasma flow rate 15 L/min | Power RF 1300 W | ||||
Auxiliary flow rate 0.2 L/min | Plasma view Axial | ||||
Nebulizer flow rate 0.6 L/min | View distance 15.0 mm | ||||
Flow Injection Programme | |||||
Step | Time (s) | Pump 1 (U/min) | Pump 2 (U/min) | Valve | Read |
Prefil | 15 | 80 | 100 | Fill | - |
1 | 10 | 80 | 100 | Fill | - |
2 | 15 | 80 | 100 | Inject | On |
Spectral Peak Processing | Time Processing | ||||
Peak Algorithm: Peak Height | Peak Algorithm: Peak Height | ||||
Points per peak: 3 points | Smooting points: 19 points | ||||
Spectral corrections: None |
|