X-ray tube output at 0.375 mA | Half value layer at 76Kv, 0.375mA | Scatter | ||||
Scan No. | Scan time (s) | mA × s | Radiation output | Thickness of Al (in mm) | Dose received (in ) | Exposure on Work station (in ) |
Scan 1 | 164 | 61.500 | 10.00 | Without filter | 10.15 | 0.40 |
Scan 2 | 159 | 59.625 | 09.80 | 0.6 | 09.80 | 0.40 |
Scan 3 | 163 | 61.125 | 09.80 | 1.2 | 09.80 | 0.35 |
Scan 4 | 165 | 61.875 | 09.60 | 1.8 | 09.00 | 0.40 |
Scan 5 | 161 | 60.375 | 10.20 | 2.4 | 08.00 | 0.40 |
Scan 6 | 163 | 61.125 | 11.00 | 3.0 | 07.00 | 0.40 |
Scan 7 | 164 | 61.500 | 09.20 | 3.6 | 06.00 | 0.35 |
Scan 8 | 163 | 61.125 | 09.80 | 4.2 | 06.00 | 0.35 |
Scan 9 | 165 | 61.875 | 10.30 | 4.8 | 05.00 | 0.40 |
Scan 10 | 158 | 59.250 | 10.00 | 5.4 | 05.00 | 0.40 |