Equipment/detector | Model | Purpose |
3D water tank | Sun Nuclear 3D SCANNERTM | All measurements except for absolute dose calibration |
1D water tank | Sun Nuclear 1D SCANNERTM | Absolute dose calibration |
Ion chamber | Sun Nuclear SNC125cTM (volume: 0.125 cm3) | Beam scans and output factors for square fields ≥ 4 × 4 cm2 |
Diode | Sun Nuclear Edge DetectorTM (volume: 0.0019 cm3) | Beam scans for square fields ≤ 3 × 3 cm2 and stereotactic cones; Output factors for square fields ≤ 4 × 4 cm2 and stereotactic cones |
Reference chamber | Sun Nuclear Reference Detector (parallel plate chamber, volume: 39 cm3) | Beam scans for square fields ≤ 3 × 3 cm2 and stereotactic cones |
Farmer-type ion chamber | Standard Imaging Exradin A12 (volume: 0.64 cm3) | Absolute dose calibration, MLC transmission and X-jaws transmission |
QA device | Sun Nuclear MapCHECK® 2 | MPPG 5.a. Tests including AAPM TG-119 plans |
Sun Nuclear SRS MapCHECK® and StereoPHANTM | Clinical SRS and SBRT plans | |
Sun Nuclear StereoPHANTM and Standard Imaging Exradin A16 (volume: 0.007 cm3) | End-to-End testing |