Measurement data | Measurement setup | Equipment/detectors | ||||
Beam modeling | ||||||
Beam scans | PDDs | Square fields (Jaws collimated)* | SSD: 100 cm | 3D water tank, ion chamber(s), diode, reference chamber | ||
Stereotactic cones† | SSD: 100 cm | 3D water tank, diode, reference chamber | ||||
Profiles | Square fields (Jaws collimated)* | SSD: 100 cm; depth: dmax, 5 cm, 10 cm, 20 cm | 3D water tank, ion chamber(s), diode, reference chamber | |||
Stereotactic cones† | SSD: 100 cm; depth: dmax, 5 cm, 10 cm | 3D water tank, diode, reference chamber | ||||
Output factors | Square fields (Jaws collimated)* | SSD: 100 cm; depth: 10 cm | 3D water tank, ion chamber, diode | |||
Stereotactic cones† | SSD: 100 cm; depth: 10 cm | 3D water tank, ion chamber, diode | ||||
X-jaws transmission | SSD: 100 cm; depth: dmax | 3D water tank, farmer-type ion chamber | ||||
MLC modeling | ||||||
MLC model parameters | x-position offset, gain and curvature; y-position gain | PDDs | Square fields (MLC collimated)‡ | SSD: 100 cm | 3D water tank, ion chamber(s), diode, reference chamber | |
Profiles | SSD: 100 cm; depth: dmax, 5 cm, 10 cm, 20 cm | |||||
Leaf tip width | Custom plan | SSD: 100 cm; depth: 10 cm | 3D water tank, diode, reference chamber | |||
Tongue and groove | Custom plan | SSD: 100 cm; depth: 10 cm | 3D water tank, diode, reference chamber | |||
Transmission | Custom plan | SSD: 100 cm; depth: dmax | 3D water tank, farmer-type ion chamber | |||
Output factors | Square fields (MLC collimated)† | SSD: 100 cm; depth: 10 cm | 3D water tank, ion chamber, diode | |||
Absolute dose calibration | ||||||
AAPM TG 51 | SSD: 100 cm; depth: 10 cm; Field size: 10 × 10 cm2 | 1D water tank, farmer-type ion chamber | ||||