Femoral component position and orientation | Difference (achieved using conventional instrumentation surgery vs. calculated using conceptual active surgical plan (SEISPT) | RMSE with Navio® Surgical System [38] | Robotic implant alignment accuracy error (median ± std. dev.) [39] |
|
Medial-lateral | 4.23 mm | NS | NS |
|
Anterior-posterior | 4.10 mm | 1.46 mm | NS | |
Superior-inferior | 6.73 mm | 0.86 mm | NS | |
Flexion-extension | 0.18˚ | 2.18˚ | 0.6˚ ± 0.5˚ | |
Varus-valgus | 3.73˚ | 0.71˚ | 0.6˚ ± 0.3˚ | |
Internal-external | 3.98˚ | 0.69˚ | 0.8˚ ± 0.5˚ | |
Tibial component position and orientation | Difference (achieved using conventional instrumentation surgery vs. calculated using conceptual active surgical plan (SEISPT) | RMSE with Navio® Surgical System [38] | Robotic implant alignment accuracy error (median ± std. dev.) [39] |
|
Medial-lateral | 3.68 mm | 1.29 mm | NS |
|
Anterior-posterior | 1.26 mm | 0.79 mm | NS | |
Superior-inferior | 7.60 mm | 0.68 mm | NS | |
Flexion-extension | 1.58˚ | 0.88˚ | 1.1˚ ± 1.6˚ | |
Varus-valgus | 2.45˚ | 0.69˚ | 0.9˚ ± 0.4˚ | |
Internal-external | 3.20˚ | 2.21˚ | NS |