Project

Infrared technology

X-ray fluorescence analysis

Neutron Activation Analysis

Prompt Gamma Neutron Activation Analysis

Simultaneous measurement of multiple elements

Yes

Atomic number > 12

Yes (Except H, C, O, etc.)

Yes

Granularity requirements

Have higher requirements

>6 mm

None

None

Influence of surrounding environment

Affected by dust, humidity, etc.

None

None

None

Real-time online analysis

Yes

Yes

Offline measurement

Yes

Measurement method

Surface

Near surface

Volume measurement

Volume measurement

Accuracy

Poor

Poor

Very high

Good

Regular correction

Yes

Yes

No

No

Radioactivity around the instrument

None

Within safe range

Within safe range

Within safe range