Linearity (N = 2)

Range (µg/L)

Slope

Intercept

R2

5 - 150

0.9983

0.0459

0.9997

Precision

%CV at 5 µg/L

%CV at 25 µg/L

%CV at 50 µg/L

%CV at 100 µg/L

Within Run (N = 5)

1.3

0.5

0.4

0.8

Between Run (N = 15)

2.1

1.7

1.6

1.0

Accuracy

% Recovery at 5 µg/L

% Recovery at 25 µg/L

% Recovery at 50 µg/L

% Recovery at 100 µg/L

Within Run (N = 5)

93.3

102.9

100.1

102.0

Between Run (N = 15)

94.8

100.9

99.0

101.0

Comparison

(N = 20)

Slope

Intercept

Bias (µg/L)

R2

0.94

1.2

4.32

0.999

Sensitivity

LOD (µg/L)

% Recovery/% CV

LOQ (µg/L)

%Recovery/% CV

0.5

121/22

1.0

102/2.5

Carry Over

% Carry Over

0.02%

System Suitability Parameters

Compound

Retention Time (min)

Theoretical Plate

Tailing Factor

Resolution Factor

5-FC

3.37

7189

1.13

26

Internal Standard

6.05

11,174

1.14