Linearity (N = 2) | Range (µg/L) | Slope | Intercept | R2 |
5 - 150 | 0.9983 | 0.0459 | 0.9997 | |
Precision | %CV at 5 µg/L | %CV at 25 µg/L | %CV at 50 µg/L | %CV at 100 µg/L |
Within Run (N = 5) | 1.3 | 0.5 | 0.4 | 0.8 |
Between Run (N = 15) | 2.1 | 1.7 | 1.6 | 1.0 |
Accuracy | % Recovery at 5 µg/L | % Recovery at 25 µg/L | % Recovery at 50 µg/L | % Recovery at 100 µg/L |
Within Run (N = 5) | 93.3 | 102.9 | 100.1 | 102.0 |
Between Run (N = 15) | 94.8 | 100.9 | 99.0 | 101.0 |
Comparison (N = 20) | Slope | Intercept | Bias (µg/L) | R2 |
0.94 | 1.2 | 4.32 | 0.999 | |
Sensitivity | LOD (µg/L) | % Recovery/% CV | LOQ (µg/L) | %Recovery/% CV |
0.5 | 121/22 | 1.0 | 102/2.5 | |
Carry Over | % Carry Over | 0.02% | ||
System Suitability Parameters | ||||
Compound | Retention Time (min) | Theoretical Plate | Tailing Factor | Resolution Factor |
5-FC | 3.37 | 7189 | 1.13 | 26 |
Internal Standard | 6.05 | 11,174 | 1.14 |