HPLC

Stationary phase

Agilent, Eclipse XDB-C18, 4.6 mm × 50 mm, 1.8 μm

Injection volume, μL

3

Detector

DAD at 250 nm

Mobile phase

A: 10 mM NH4Ac(aq) 90%/MeOH 10%

B: 10 mM NH4Ac(aq) 20%/MeOH 80%

Degassed in ultrasonic bath for 20 min

Gradient program

0 - 3 min 50% B, isocratic

3 - 12 min up to 100% B, held for 5 min

17 - 23 min down to 5% B, held for 5 min to wash out salts from biomatrices

28 - 31 min up to 50% B, held for 4 min to re-condition column

Flow rate, mL・min−1

0.5

Electrospray mass spectrometry

Temperature (˚C)

400

Detector polarity

Positive ion

Scanning mass range

100 - 1000

Resolution, Q1 and Q3

Unit

Nebulizer gas (p.s.i.)

40

Curtain gas (p.s.i.)

10

Collision gas (p.s.i.)

Medium or high for EMS and EPI scan modesa, respectively

Ion spray voltage (V)

5000

Declustering potential (V)

40