| HPLC | |
| Stationary phase | Agilent, Eclipse XDB-C18, 4.6 mm × 50 mm, 1.8 μm |
| Injection volume, μL | 3 |
| Detector | DAD at 250 nm |
| Mobile phase | A: 10 mM NH4Ac(aq) 90%/MeOH 10% B: 10 mM NH4Ac(aq) 20%/MeOH 80% Degassed in ultrasonic bath for 20 min |
| Gradient program | 0 - 3 min 50% B, isocratic 3 - 12 min up to 100% B, held for 5 min 17 - 23 min down to 5% B, held for 5 min to wash out salts from biomatrices 28 - 31 min up to 50% B, held for 4 min to re-condition column |
| Flow rate, mL・min−1 | 0.5 |
| Electrospray mass spectrometry | |
| Temperature (˚C) | 400 |
| Detector polarity | Positive ion |
| Scanning mass range | 100 - 1000 |
| Resolution, Q1 and Q3 | Unit |
| Nebulizer gas (p.s.i.) | 40 |
| Curtain gas (p.s.i.) | 10 |
| Collision gas (p.s.i.) | Medium or high for EMS and EPI scan modesa, respectively |
| Ion spray voltage (V) | 5000 |
| Declustering potential (V) | 40 |