Parameter | Value | Column 1 | Column 2 |
Instrument settings | Agilent 5977B MSD conditions | ||
Vial pressurization gas | Helium | Acquisition mode | Scan (20 m/z - 450 m/z)) |
Loop size | 1 mL | Solvent delay | 1.73 minutes |
Keyboard lock | ON | Tune file | etune.u |
Transfer line type | Fused Silica | EM Setting mode | Gain = 1 |
Transfer line diameter | 0.53 mm | Scan speed | 781 [N = 3] |
System configuration | Scan parameters | ||
Carrier control | GC Instrument | Threshold | 50 |
Oven temperature | 140˚C | MSD Source temperature | 280˚C |
Loop temperature | 140˚C | MSD Quadrupole temperature | 150˚C |
Transfer line temperature | 160˚C | ||
Vial equilibration | 5.00 minutes | ||
Injection duration | 0.25 minutes | ||
GC cycle time | 30 minutes | ||
Vial size | 20 mL | ||
Vial shaking | Level 2, 25 shakes/min | ||
Fill mode | Default | ||
Fill pressure | 15 psi | ||
Loop fill mode | Custom | ||
Loop ramp rate | 20 psi/min | ||
Loop final pressure | 10 psi | ||
Loop equilibration time | 0.05 minutes | ||
Carrier control mode | GC controls carrier | ||
Extraction mode | Single extraction | ||
Vent after extraction | ON | ||
Post injection purge | 100 mL/min for 1 minute |
| |
Acceptable leak check | Default, 0.2 mL/min | ||
Agilent 7890B GC conditions | |||
GC oven temperature | 40˚C | ||
Hold time | 5 minutes | ||
Oven program | 15˚C/min to 160˚C, hold 0 minutes, then 35˚C/min to 300˚C, hold 7 minutes |
| |
Equilibration time | 0.1 minute | ||
Max temperature | 320˚C | ||
Column flow | 1.2 mL/min | ||
Font S/SL inlet He | Split | ||
Column specifications | |||
Catalog number | Restek 13868 | ||
Product name | Rxi-624Sil MS | ||
Size | 30 meter × 0.25 mm ID × 1.4 µm df |
| |
Max temperature | 320˚C |