N

Measurand (measurable value)

Object of measuring

Method

Measuring range

Uncertainty

1

Absorbed dose rate to water

Calibrated field from linear accelerator

Measurements with chemical dosimeters

1 × 102 ... 1 × 105 Gy∙s−1

2.5%

2

Absorbed dose to water

Chemical dosimeters

Irradiation in calibrated field from linear accelerator (high level)

1 × 101 ... 1 × 106 Gy

2.5%

3

Absorbed dose rate to water

Calibrated field from Sr-90/Y-90 radionuclide sources

Measurements with chemical dosimeters

3 × 10−1 ... 1 × 102 Gy∙s−1

2.5%

4

Absorbed dose to water

Chemical dosimeters

Irradiation in calibrated field from Co-60 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.2%

5

Absorbed dose to water

Chemical dosimeters

Irradiation in calibrated field from Cs-137 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.2%

6

Absorbed dose to graphite

Chemical dosimeters

Irradiation in calibrated field from Co-60 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.2%

7

Absorbed dose to graphite

Chemical dosimeters

Irradiation in calibrated field from Cs-137 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.2%

8

Absorbed dose to aluminium

Chemical dosimeters

Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.5%

9

Absorbed dose to graphite

Chemical dosimeters

Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.5%

10

Absorbed dose to PMMA

Chemical dosimeters

Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

2.5%

11

Absorbed dose to polystyrene

Chemical dosimeters

Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

2.5%

12

Absorbed dose to polycarbonate

Chemical dosimeters

Irradiation in calibrated field from Sr-90/Y-90 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

2.5%

13

Absorbed dose to polystyrene

Chemical dosimeters

Irradiation in calibrated field from Co-60 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.2%

14

Absorbed dose to polystyrene

Chemical dosimeters

Irradiation in calibrated field from Cs-137 radionuclide sources (high level)

1 × 101 ... 1 × 106 Gy

1.2%