BEFORE the corrections for artifacts (60Co beam)

Depth 1

Thickness (mm)

Measured dose for a field size (cGy)

% deviation between measured and calculated

7 × 7 cm2

10 × 10 cm2

15 × 15 cm2

7 × 7 cm2

10 × 10 cm2

15 × 15 cm2

0

199.630

200.000

200.530

−0.18

0.00

0.26

4

207.270

207.460

207.060

3.64

3.73

3.53

8

207.330

209.040

208.130

3.66

4.52

4.06

12

204.910

206.280

205.890

2.45

3.14

2.94

8 (air gap)

212.980

212.700

209.920

6.49

6.35

4.96

Depth 2

Thickness (mm)

Measured dose for a field size (cGy)

% deviation between measured and calculated

7 × 7 cm2

10 × 10 cm2

15 × 15 cm2

7 × 7 cm2

10 × 10 cm2

15 × 15 cm2

0

199.630

200.000

200.530

−0.18

0.00

0.26

4

204.090

204.800

204.790

2.04

2.40

2.40

8

205.250

206.680

206.700

2.62

3.34

3.35

12

199.020

201.420

201.470

−0.49

0.71

0.74

8 (air gap)

205.350

206.270

206.450

2.68

3.13

3.23