| Parameter |
S-Epichlorohydrin |
| Linearity |
|
| Correlation coefficient |
0.9998 |
| Slope |
137034.3 |
| Y-Intercept |
0.0 |
| %Y-Intercept |
0.22 |
| Accuracy (% Recovery) |
|
| LOQ (n = 3) |
98.2 |
| 50% (n = 3) |
102.8 |
| 100% (n = 3) |
101.5 |
| 150% (n = 3) |
99.8 |
| Precision (% RSD) |
|
| LOQ (n = 6) |
3.29 |
| 50% (n = 6) |
1.98 |
| 100% (n = 6) |
1.08 |
| 150% (n = 6) |
0.98 |
| Ruggedness; Different day and analyst (%RSD) |
|
| 100% (n = 6) |
2.16 |
| Robustness (%RSD) |
|
| Actual flow 25 psi |
0.80 |
| Different flow 20 psi |
1.34 |
| Different flow 30 psi |
1.09 |
| Different Column Temperature 45˚C |
2.12 |
| Different Column Temperature 55˚C |
1.98 |
| Limit of Detection (Concentration in µg/mL) |
0.001 |
| Limit of Quantification (Concentration in µg/mL) |
0.004 |