Parameter | S-Epichlorohydrin |
Linearity | |
Correlation coefficient | 0.9998 |
Slope | 137034.3 |
Y-Intercept | 0.0 |
%Y-Intercept | 0.22 |
Accuracy (% Recovery) | |
LOQ (n = 3) | 98.2 |
50% (n = 3) | 102.8 |
100% (n = 3) | 101.5 |
150% (n = 3) | 99.8 |
Precision (% RSD) | |
LOQ (n = 6) | 3.29 |
50% (n = 6) | 1.98 |
100% (n = 6) | 1.08 |
150% (n = 6) | 0.98 |
Ruggedness; Different day and analyst (%RSD) | |
100% (n = 6) | 2.16 |
Robustness (%RSD) | |
Actual flow 25 psi | 0.80 |
Different flow 20 psi | 1.34 |
Different flow 30 psi | 1.09 |
Different Column Temperature 45˚C | 2.12 |
Different Column Temperature 55˚C | 1.98 |
Limit of Detection (Concentration in µg/mL) | 0.001 |
Limit of Quantification (Concentration in µg/mL) | 0.004 |