Urine validation results

Methamphetamine

Amphetamine

Cathinone

Methcathinone

S(+)

R(−)

S(+)

R(−)

R(+)

S(−)

R(+)

S(−)

Extraction recovery

94%

98%

86%

85%

70%

70%

85%

86%

Matrix effect

114%

106%

108%

109%

111%

113%

118%

117%

Precision RSD (intra-day) (6)

2.7%

3.7%

3.4%

2.3%

4.0%

4.4%

3.8%

3.2%

Precision RSD (inter-day) (6 + 6 + 6)

2.8%

4.0%

2.8%

3.2%

3.6%

3.7%

4.3%

3.3%

Lower limit of detection (pg/mL)

20

20

200

200

250

250

20

20

Lower limit of confirmation (pg/mL)

50

50

500

500

500

500

50

50

Plasma validation results

Methamphetamine

Amphetamine

Cathinone

Methcathinone

S(+)

R(−)

S(+)

R(−)

R(+)

S(−)

R(+)

S(−)

Extraction recovery

100%

97%

95%

98%

84%

84%

100%

94%

Matrix effect

109%

112%

118%

117%

98%

112%

113%

123%

Precision RSD (intra-day) (6)

3.7%

2.2%

4.3%

4.7%

3.5%

2.9%

3.8%

3.7%

Precision RSD (inter-day) (6 + 6 + 6)

4.5%

4.6%

4.8%

4.6%

3.9%

2.9%

4.3%

3.2%

Lower limit of detection (pg/mL)

2

2

20

20

10

10

2

2

Lower limit of confirmation (pg/mL)

5

5

50

50

25

25

5

5